2 source scpi command reference, Source scpi command reference -18 – AMETEK M130 User Manual
Page 68
IEEE 488.2 and SCPI Command Operation
Sorensen Ethernet Option
3-18
M130/131 Programming Manual
3.8.2
SOURCE
SCPI
C
OMMAND
R
EFERENCE
Command
Description
SCPI
SOURce[n]
Source subsystem. n = 1-31. The default channel
is 1.
C
:CURRent
Sets the output current in amps (default) or in
milliamps.
C
[:LEVel]
Sets the output current in amps (default) or in
milliamps.
C
[:IMMediate]
Sets the output current in amps (default) or in
milliamps.
C
[:AMPLitude] <0+NRf>
Sets the output current in amps (default) or in
milliamps.
C
:TRIGgered
Sets the value of the output current to be
implemented by the trigger.
C
:CLEar
Clears the trigger mode.
[:AMPLitude] <0+NRf>
Sets the value of the output current to be
implemented by the trigger.
C
:LIMit
Sets an upper soft limit on the programmed
output current for the supply.
C
[:AMPLitude] <0+NRf>
Sets an upper soft limit on the programmed
output current for the supply.
C
:RAMP <0+NRf> <0+NRf>
Sets the output current to ramp from the present
value to the specified value (first argument) in
the specified time (second argument). See Ramp
Function description below.
N
:ABORt
Aborts ramping and clears trigger mode.
:ALL?
Returns the ramping status of all channels.
:TRIGgered <0+NRf>
<0+NRf>
Sets the output current to ramp from the present
value to the specified value (first argument) in
the specified time (second argument) upon the
trigger command. See Ramp description below.
N
:ONLine?
Returns the integer value 1(ONLINE) or 0
(OFFLINE) of the channel online status.
N
:STATus
:BLOCk?
Returns the block of data critical to the status of
the channel:
1) channel number
2) online status
3) status flags register (see table below)
4) status register (see sect. 5.2.2)
5) accumulated status (see sect. 5.2.2)
6) fault mask register (see sect. 5.2.2)
7) fault register (see sect. 5.2.2)
8) error register (see sect. 5.2.3)
9) model serial number
10) model voltage
11) model current
12) model over-voltage
N