5 readback resolution, 6 readback accuracy, Readback resolution -5 – AMETEK M130 User Manual
Page 17: Readback accuracy -5

Sorensen Ethernet Option
Features, Functions and Specifications
M130/M131 Programming Manual
1-5
1.3.5
R
EADBACK
R
ESOLUTION
DLM600
DCS
Voltage
0.002% of full scale
0.02% of full scale
Current
0.002% of full scale
0.02% of full scale
1.3.6
R
EADBACK
A
CCURACY
DLM600
DCS
Voltage
± (0.1% of full scale output voltage)
± (0.1% + 0.15% of
maximum output voltage)
Current*
± (0.25% of full scale output current)
*
± (0.1% + 0.4% of
maximum output current)
*
After 30 minutes operation with fixed line, load, and temperature.
Note: Refer to the applicable power supply manual (DLM600 or DCS) for effects of
line regulation, load regulation, and temperature on accuracy specifications.