IEEE 488.2 and SCPI Command Operation

Sorensen Ethernet Option

3-12

M130/131 Programming Manual

:VOLTage

:DAC <0+NR1>

:FIVEPOINT <1|2|3|4|5> <0+NRf>

:FIVEPOINT?

:GAIN

:OFFSet

:PROTection

:CALCulate

:DAC <0+NR1>

:GAIN

:OFFSet

:STORe

:UNLock

3.5.2

CALIBRATION

SCPI

C

OMMAND

R

EFERENCE

Command

Description

SCPI

CALibrate[n]

Calibration subsystem. n = 1-31. The
default channel is 1.

C

:DATA



Sets the values of the ten floating point
calibration constants:
1) output voltage DAC gain
2) output voltage DAC offset
3) output current DAC gain
4) output current DAC offset
5) output voltage protection DAC gain
6) output voltage protection DAC offset
7) voltage measurement ADC gain
8) voltage measurement ADC offset
9) current measurement ADC gain
10) current measurement ADC offset
Values are separated by space or comma.

N

:INITial

N

:CURRent

Sets the power-on default value of current.

N

:VOLTage

N

[:AMPLitude]

Sets the power-on default voltage.

N

:PROTection

Sets the power-on default value of the
overvoltage protection.

N

:LOCK

Disables access to the non-volatile memory.
Prevents attempts to store calibration
values.

N

:MEASure

N

:CURRent

:ADC?

Returns the integer value of the A/D for the
current measurement.

N

:CALCulate

Calculates the value of the gain and offset
for current measurements.

N

This manual is related to the following products: