2 calibration scpi command reference, Calibration scpi command reference -12 – AMETEK M130 User Manual
Page 62

IEEE 488.2 and SCPI Command Operation
Sorensen Ethernet Option
3-12
M130/131 Programming Manual
:VOLTage
:DAC <0+NR1>
:FIVEPOINT <1|2|3|4|5> <0+NRf>
:FIVEPOINT?
:GAIN
:OFFSet
:PROTection
:CALCulate
:DAC <0+NR1>
:GAIN
:OFFSet
:STORe
:UNLock
3.5.2
CALIBRATION
SCPI
C
OMMAND
R
EFERENCE
Command
Description
SCPI
CALibrate[n]
Calibration subsystem. n = 1-31. The
default channel is 1.
C
:DATA
Sets the values of the ten floating point
calibration constants:
1) output voltage DAC gain
2) output voltage DAC offset
3) output current DAC gain
4) output current DAC offset
5) output voltage protection DAC gain
6) output voltage protection DAC offset
7) voltage measurement ADC gain
8) voltage measurement ADC offset
9) current measurement ADC gain
10) current measurement ADC offset
Values are separated by space or comma.
N
:INITial
N
:CURRent
Sets the power-on default value of current.
N
:VOLTage
N
[:AMPLitude]
Sets the power-on default voltage.
N
:PROTection
Sets the power-on default value of the
overvoltage protection.
N
:LOCK
Disables access to the non-volatile memory.
Prevents attempts to store calibration
values.
N
:MEASure
N
:CURRent
:ADC?
Returns the integer value of the A/D for the
current measurement.
N
:CALCulate
Calculates the value of the gain and offset
for current measurements.
N