AMETEK M130 User Manual
Page 64
IEEE 488.2 and SCPI Command Operation
Sorensen Ethernet Option
3-14
M130/131 Programming Manual
Command
Description
SCPI
:OFFSet
Sets the value of the offset for the output
voltage.
N
:PROTection
N
:CALCulate
Calculates the value of the gain and offset
for output overvoltage protection. This takes
more than 30 seconds to complete. Use
*ESE 1 and a serial poll to detect the
completed operation.
N
:DAC <0+NR1>
Sets the output of the output overvoltage
protection D/A converter. The value range
is 0 – 4095 for 12-bit operations; 0-65535
for 16-bit.
N
:GAIN
Sets the value of the gain for the output
overvoltage protection.
N
:OFFSet
Sets the value of the offset for the output
overvoltage protection.
N
:STORe
Stores the calibration constants in non-
volatile memory.
N
:UNLock
Sets the non-volatile memory available to
store calibration constants. The access
string is “6867”.
N