Checkline TI-MINIMAX User Manual
Page 86

Dakota Ultrasonics
80
Note: Since this is the 1
st
verification following the initial measurement of
the
Probe Zero Cal Bars, the cursor should be located on a cal standard in
column B.
20) If the elongation is +/- .0001” and “In Spec”, Press the
ENTER key to
save the current reading in the highlighted cell location, storing all the
current probe zero verification data.
Repeat steps 6 - 31 for the other
probe zero calibration standards. If the probe zero is “Out of
Spec”, proceed to the next step.
21)
Assuming the elongation is “Out of Spec”, Press the MEAS key once to
activate the measure menu items. Press the
MEAS key multiple times
to move right and the
ESC key multiple times to move left, until the
TEMP cell is highlighted.
22) Use the
LEFT and RIGHT arrow keys to adjust the temperature value
util the elongation is 0.0000”, or “In Spec” at +/- 0.0001”.
23) . Press the
MEAS key multiple times to move right and the ESC key
multiple times to move left, until the
LOC cell is highlighted.
24) Press the
ENTER key to display the Group View Box. Notice the
elongation value is now “In Spec”.
25) Press the
ENTER key to save the current reading in the highlighted cell
location, storing all current probe zero verification data.
26)
Repeat steps 6 - 31 for the other probe zero calibration standards.
Once again, If the probe zero is “Out of Spec”, repeat steps 32 – 36.