Checkline TI-MINIMAX User Manual
Page 76
Dakota Ultrasonics
70
1) With the menu items still active from the previous section, press the
MENU key multiple times to tab right and the ESC key multiple times to
tab left until the
CAL menu is highlighted and displaying the submenu
items.
2) Use the
UP and DOWN arrow keys to scroll through the sub menu items
until
ZERO MODE is highlighted.
3) Use the
LEFT and RIGHT arrow keys to scroll the zero mode options
until the
TWO POINT option is displa yed.
Note: The two-point option allows the user to use two separate points for
the probe zero calibration. Note: The one point option can optionally be
selected if only one cal bar will be used. Although the procedures below
are described with the two-point option in mind, they will also work in the
case of a one-point calibration.
4) Use the
UP and DOWN arrow keys to scroll through the sub menu items
until
MEASURE ZERO is highlighted.
5) Press the
ENTER key to display the Two Point Zero list box.
6) Press the
UP and DOWN arrow keys to scroll through the Two Point
Zero list until PHY LEN 1 is highlighted.
Note: PHY LEN 1 is the actual physical length of the first calibration
standard.
7) Press the
ENTER key to display the Digits Edit Box.
8) Press the
UP and DOWN arrow keys to scroll the highlighted value.