Checkline TI-MINIMAX User Manual
Page 77

MINI-MAX Ultrasonic Bolt Tension Monitor
9) Press the
LEFT and RIGHT arrow keys to scroll the digit locations.
10) Repeat steps
8 & 9 until the PHY LEN 1 value is correctly displayed.
11) Press the
OK key to return to the Two Point Zero list, or ESC to cancel
entering the
PHY LEN 1.
12) Press the
UP and DOWN arrow keys to scroll through the Two Point
Zero list until PHY LEN 2 is highlighted.
Note: PHY LEN 2 is the actual physical length of the second calibration
standard.
13) Press the
ENTER key to display the Digits Edit Box.
14) Press the
UP and DOWN arrow keys to scroll the highlighted value.
15) Press the
LEFT and RIGHT arrow keys to scroll the digit locations.
16) Repeat steps
14 & 15 until the PHY LEN 2 value is correctly displayed.
17) Press the
OK key to return to the Two Point Zero list, or ESC to cancel
entering the
PHY LEN 2.
18) Press the
UP and DOWN arrow keys to scroll through the Two Point
Zero list until VELOCITY is highlighted.
Note: This refers the velocity of the probe zero cal bars.
19) Press the
ENTER key to display the Digits Edit Box.
20) Press the
UP and DOWN arrow keys to scroll the highlighted value.
21) Press the
LEFT and RIGHT arrow keys to scroll the digit locations.
22) Repeat steps
20 & 21 until the VELOCITY value is correctly displayed.
23) Press the
OK key to return to the Two Point Zero list, or ESC to cancel
entering the
VELOCITY.