MINI-MAX Ultrasonic Bolt Tension Monitor
1) Press the
MEAS key once to activate measure menu items. Press the
MEAS key multiple times to move right and the ESC key multiple times
to move left until the
LOC cell is highlighted.
2) Press the
ENTER key to display the Group View Box.
3) Press the
UP, DOWN, LEFT, and RIGHT arrow keys to scroll the target
cell cursor to the desired storage location.
Note: The first column must be used to store reference lengths.
4) Press the
ENTER key to save the current reading in the highlighted cell
location.
5) Abort the
Grid View Box by pressing the MEAS key at any time.
6) Proceed to the next section to perform and store the first
Probe Zero
verification. DO NOT REMOVE THE TRANSDUCER FROM THE
BOLT.
The next two sections outline the procedures of how to perform a probe zero
calibration by calculating a zero value, or alternatively adjusting the temperature
value to match the probe zero value. When should I use the
Calc Zero feature
versus adjusting the
Temperature, and why?
The only time the
Calc Zero feature should be used is when a different MINI-
MAX, transducer, or transducer cable is being used, other than the one initially
used to obtain the probe zero data for the probe zero calibration bolt. In this
case, calculating the probe zero value forces the new hardware to match with the