Checkline TI-MINIMAX User Manual
Page 62

Dakota Ultrasonics
56
10) Repeat steps
8 & 9 until the PHYSICAL LEN value is correctly
displayed.
11) Press the
OK key to return to the One Point Zero List, or ESC to
cancel entering the
PHYSICAL LEN value.
12) Use the
UP and DOWN arrow keys to scroll through the One Point
Zero List Items until VELOCITY is highlighted.
13) Press the
ENTER key to display the Digits Edit Box.
Note: Enter the velocity value of the bolt that was documented in the
previous section. Reminder: The value was written down and scribed on
the side of the bolt. It is the velocity used for the probe zero bolt.
14) Press the
UP and DOWN arrow keys to scroll the highlighted value.
15) Press the
LEFT and RIGHT arrow keys to scroll the digit locations.
16) Repeat steps
14 & 15 until the VELOCITY value is correctly displayed.
17) Press the
OK key to return to the One Point Zero List, or ESC to
cancel entering the
VELOCITY value.
18) Use the
UP and DOWN arrow keys to scroll through the One Point
Zero List Items until TEMP COEF is highlighted.