user does not increase it at that time, when the fastener is put under a load, the
same peak jumps can occur. Even though the threshold can be adjusted at a
later time to correct the peak jump, errors can be overlooked during the
measurement process. Therefore, it’s always best to make these adjustments
when initially measuring the reference lengths.
The procedures to adjust the
Threshold are outlined below:
Adjusting the Threshold using the Hot Menus
1) Press the
MEAS key once to activate measure menu items. Press the
MEAS key multiple times to move right and the ESC key multiple times
to move left, until the
THR (threshold) cell is highlighted.
2) Press the
UP, DOWN, LEFT, and RIGHT arrow keys to scroll the
highlighted value.
3)
Alternatively, press the ENTER key to display the Digits Edit Box.
4) Press the
UP and DOWN arrow keys to scroll the highlighted value.
5) Press the
LEFT and RIGHT arrow keys to scroll the digit locations.
6) Repeat steps
4 & 5 until the THR value is correctly displayed.
7) Press the
OK key to return to the measurement screen, or ESC to
cancel entering the
THR.