Checkline TI-MINIMAX User Manual
Page 57

MINI-MAX Ultrasonic Bolt Tension Monitor
12) Use the
UP and DOWN arrow keys to scroll through the sub menu
items until
QUANTITY is highlighted.
13) Use the
LEFT and RIGHT arrow keys to scroll the quantity options until
the
ELONG option is displayed.
14) Press the
MENU key multiple times to tab right and the ESC key
multiple times to tab left until the
AUTO menu is highlighted and
displaying the submenu items.
15) Use the
UP and DOWN arrow keys to scroll through the sub menu
items until
APROX LEN is highlighted.
Note: This is the approximate physical length of the fastener or bolt. This
tells the MINI-MAX where to turn on the receiver and look for an echo. The
area scanned will be +/- 5% from the approximate length.
16) Press the
ENTER key to display the Digits Edit Box.
17) Press the
UP and DOWN arrow keys to scroll the highlighted value.
18) Press the
LEFT and RIGHT arrow keys to scroll the digit locations.
19) Repeat steps
16 & 17 until the APPROX LEN value is correctly
displayed.