MINI-MAX Ultrasonic Bolt Tension Monitor
3) Record the ultrasonic length “L
0
” of the sample bolt at the above
minimum temperature - 48
0
F.
4)
Repeat step 1 - 3 at the target temperatures (T
1
through T
5
) – 48, 68,
88, and 108
0
F (other similar range with 5 temperatures), measuring the
ultrasonic lengths (L
1
through L
5
) respectively. This process must be
done for each of the sample bolts in the experiment.
5) Use the equation below to calculate a linear regression, using the
temperatures and lengths recorded for all the sample bolts to determine
the temperature factor.
( )( )
( )
6
2
2
10
∗
−
−
=
∑
∑
∑
∑
∑
n
x
x
n
y
x
xy
eFactor
Temperatur
x = temperature, y = change in reference length, n = number of sample points per bolt
See also other documents in the category Checkline Sensors:
- Series-2 (16 pages)
- Series-3 (22 pages)
- Series-4 (26 pages)
- Series-5 (34 pages)
- FGV-XY (19 pages)
- DS2 (6 pages)
- FGE (8 pages)
- Series-5I (32 pages)
- Series-3I (24 pages)
- MG (2 pages)
- FGE-XY (6 pages)
- FB (4 pages)
- ES10 (4 pages)
- FGS-100H (2 pages)
- FGS-250W (2 pages)
- TSB100 (10 pages)
- ES30 (6 pages)
- ES05 (4 pages)
- TSA750 (10 pages)
- TSF (6 pages)
- ESM301 (27 pages)
- MX2 (8 pages)
- FGS-100PV (20 pages)
- FGS-220VC (33 pages)
- MX-500 (4 pages)
- ESMH (6 pages)
- FGS-100PX (20 pages)
- ZP (10 pages)
- WT-100 (2 pages)
- AWS-4050 (14 pages)
- TI-25LT (9 pages)
- TI-25S (11 pages)
- TI-25M (13 pages)
- TI-25DL (19 pages)
- TI-25M-MMX (15 pages)
- TI-25DL-MMX (17 pages)
- TI-007 (9 pages)
- TI-007DL (11 pages)
- TI-CMX (29 pages)
- TI-CMXDL (35 pages)
- TI-CMXDLP (54 pages)
- TI-MVX (42 pages)
- TI-UMX2 (29 pages)
- TI-25P (11 pages)
- TI-44N (11 pages)