Setting up the mini-max for measurement – Checkline TI-MINIMAX User Manual
Page 78

Dakota Ultrasonics
72
24) Press the
UP and DOWN arrow keys to scroll through the Two Point
Zero list until TEMP COEF is highlighted.
Note: This refers to the temperature coefficient of the probe zero cal bars.
25) Press the
ENTER key to display the Digits Edit Box.
26) Press the
UP and DOWN arrow keys to scroll the highlighted value.
27) Press the
LEFT and RIGHT arrow keys to scroll the digit locations.
28) Repeat steps
20 & 21 until the TEMP COEF value is correctly
displayed.
29) Press the
OK key to return to the Two Point Zero list, or ESC to cancel
entering the
TEMP COEF.
30) Press the
ESC key once to leave the Two Point Zero list.
31) Proceed to the next section.
Before the probe zero is calculated the zero points of the cal bars need to be
ultrasonically measured and documented. This procedure is demonstrated in the
next section.
Setting up the MINI-MAX for Measurement
1) With the menu items still active from the previous section, press the
MENU key multiple times to tab right and the ESC key multiple times to
tab left until the
GEOM menu is highlighted and displaying the submenu
items.