6.1.2 Retrieving Diagnostic Data
Table 6-2 also lists the diagnostic query commands associated with each of the diagnostic codes. When a diagnos-
tic code is generated, the related query command can be used to retrieve additional information about the error.
Using the query command for a given error also clears the diagnostic flag for that error. Once cleared, that data
is no longer available.
6.2
Diagnostic Commands
Table 6-3 summarizes the diagnostic commands available for the iQUBE
2
, including query commands, parameter-
setting commands, and the values that can be assigned by the parameter-setting commands. The diagnostic com-
mands are described in detail in the following sections.
6.2.1 System Diagnostic Commands
System diagnostic commands consist of query commands only; there are no parameter-setting commands. System
diagnostics are always enabled.
For data formats that include the status token, some system diagnostic errors also set the status to
I
(invalid).
Power Supply Error,
less than 6 VDC or greater than 12 VDC).
Error Type
Query Command
Parameter Setting
Commands
Parameter Values
Default
Value Range
P
Power supply
DIA.PS
—
—
—
S
Secondary
communication
DIA.MSCONNECT
—
—
—
E
Excitation
DIA.PSEXC
—
—
—
C
Cell connection
DIA.CELLCONNECT
—
—
—
V
Overloaded cell
DIA.OVERLOAD
SCs.DIA.OVERLOAD
100
–100 – 100
(percent of cell)
U
Underloaded cell
DIA.UNDERLOAD
SCs.DIA.UNDERLOAD
10
–100 – 100
(percent of cell)
R
Zero reference
DIA.ZREF
SCs.DIA.ZREF
OFF
OFF, ON
SCs.DIA.ZREF.RANGE
2
0 – 100 (percent of scale)
SCs.DIA.ZREF.THRESH
10
–100 – 100 (percent of
scale)
SCs.DIA.ZREF.TIME
10
1 – 60 (seconds)
D
Drift
DIA.DRIFT
SCs.DIA.DRIFT
OFF
OFF, ON
SCs.DIA.DRIFT.RANGE
2
0 – 100.0
(percent of scale)
SCs.DIA.DRIFT.THRESH
2
0 – 100.0
(percent of scale)
SCs.DIA.DRIFT.TIME
10
1 – 300 (seconds)
N
Peak noise
DIA.NOISE
SCs.DIA.NOISE
OFF
OFF, ON
DIA.NOISE.THRESH
10.00
0.01 – 1000 (
V)
SCs.DIA.NOISE.TIME
10
1 – 300 (seconds)
L
Unbalanced cell
DIA.UNBAL
SCs.DIA.UNBAL
OFF
OFF, ON
SCs.DIA.UNBAL.RANGE
5
5 – 75 (percent of scale)
SCs.DIA.UNBAL.THRESH
10
0 – 50 (percent of scale)
—
All
DIA.FLAGS
—
—
—
—
All
DIA.CLEAR
—
—
—
Table 6-3. Diagnostic Command Summary