Calibration
33
4.0 Calibration
Calibration types depend on how many load cells are connected to iQUBE
2
. See Table 4-1. If only one load cell is
connected, only 2-point calibration will be available.
The iQUBE
2
can be calibrated by using the Virtui
2
Configuration Utility
,
Revolution III, 920i, or serial commands.
No minimum test weight values are required; however, at least 10% of the scale capacity is recommended.
4.1
920i Calibration
Verify that the iQUBE
2
is connected to the 920i
with power applied. Use the 920i
setup switch to make sure the
indicator is in setup mode.
Do not move the
iQUBE
2
SW1 switch. It should remain in the OFF position (see Figure on page 6 and Figure 3-
1 on page 19). If
iQUBE
2
is not connected, menus may run very slowly. Configuration must be downloaded prior
to calibrating.
Do not mix calibration types (section, corner, 2-point). Span adjustments must be made in appropriate
calibration type.
Figure 4-1. 920i CALIB Submenu
Calibration Type
Description
2-POINT
Traditional span calibration. No trimming will be done.
CORNER
Used when weight will be placed over each load cell.
SECTION
Used when load cells are paired with each other and calibrated in those pairs.
THEORETICAL
Mathematically calibrates the scale using full scale capacity and load cell sensitivities.
Table 4-1. Calibration Types
Note
Important
DIAGNOS
GRADS
FORMAT
CALIB
…
…
SCALES
SERIAL
FEATURE
PFORMT
SETPTS
DIG I/O
VERS
SCALE x
WZERO
WSPAN
WLIN
WVAL
CALMATCH
CALSPAN
WLIN
POINT 1
POINT 2
weight_value
span value
count
WGT1
CAL1
Same as
Point 1
Same as Point 1
Diagnostics
Screen
For
Trimming
CALZERO
Last Zero or
Temp Zero
SECTION
2-POINT
POINT 3
POINT 4
POINT 5
POINT 1
POINT 2
CORNER
Same as
SECTION
Theoretical
Calibration
THEORET
POINT 4
POINT 5
POINT 3
Theoretical
Calibration
deadload
count value
weight_value
span value
count
Deadload
adjustment
REZERO
WVAL
THEORET
REZERO
Deadload
adjustment