Calibration
37
Figure 4-3. Two-Point Calibration wizard
4. Select the calibration type (see Figure 4-3), then click
Next
.
5. Enter the amount of test weight being used.
6. If applicable, select the Chains or Hooks are used during calibration check box and click
Next
.
7. Remove all weight from the scale and click
Calibrate Zero
. If you want to use the zero acquired from the last
press of the Zero button, click
Use Last Zero
.
The status bar (lower left corner) of the window displays Transmitting... until calibration is finished. It then
displays Zero Calibration Complete.
8. Click
Next
.
9. Add the test weight to the scale and click
Calibrate Span
.
10. If linear points were selected in Step 5, enter the applied weight in the text box and click
Measure
.
The Status will display
Success
when complete.
Figure 4-4. Two-Point Calibration wizard
11. Click
Next
.
12. If chains or hooks are being used, remove them (along with all other weight) from the scale and click
Re-Zero
Calibrate
. Click
Next
.
13. Calibration is complete. Click
Finish
to commit the calibration values and send them to iQUBE
2
.
Click
Cancel
if you want to abort. This will require a re-calibration at a later time.
4.2.2 Corner Match
This option allows you to calibrate and trim cells based on weight placed on individual load cells.
1. From the
Scales
section, select the scale to be calibrated.
2. From the
General
tab, select
Corner Match
from the
Calibration Type
pull-down menu.
3. Click the
Calibrate
button.
The Corner Match Calibration wizard appears.
Note