Rice Lake iQUBE2 Digital Diagnostic Junction Box User Manual
Page 48

44
iQUBE
2
Installation Manual
Command
c = load cell number
Description
Values
LCc.ID
Load cell ID
Load Cell A1, load_cell_ID (16 characters,
maximum)
LCc.SRLNUM
Load cell serial number
(16 characters, maximum)
LCc.CAPACITY
Load cell capacity
0-1,000,000
LCc.SENS
Load cell sensitivity
0.0001–9.9999 (mV/V)
LCc.TRIM
Load cell trim factor
0–5.0000
LCc.ZERO
Load cell zero counts value
0–16777215
LCc.COUNTS
Read A/D count
—
LCc.CDMV
Read corrected deadload mV value
—
LCc.RMV
Read raw mV value
—
LCc.CMV
Read corrected mV value
—
LCc.WEIGHT
Read calibrated weight value
—
LCx.SAMPLEn
Uploads/Downloads calibration values for the
given (n)
Cal-Match
calibration step (
-16777215–16777215
LCx.SAMPLE
Uploads/Downloads calibration values for all
calibration steps
-16777215–16777215
Table 5-6.
iQUBE
2
Load Cell Commands
Command
s = scale number
Description
Values
SCs.CALTYPE
Calibration type
SECMATCH, CORNERMATCH, 2-
POINT
Calibration commands for CALTYPE = SECMATCH, CORNERMATCH
SCs.CAL0
Zero calibration
—
SCs.CALn
Calibration step n
Assign calibration step number (1–
16, maximum) for each load cell in
CORNERMATCH calibration, for
each section pair (1–8, maximum) in
SECTIONMATCH calibration
SCs.NORM
Normalize platform
Final calibration step after
completing SCs.CAL0 through
SCs.CALn
Calibration commands for CALTYPE = SINGLE, MULTI
SCs.WZERO
Zero calibration
—
SCs.WVAL
Platform test weight value
0–9999999
SCs.WSPAN
Platform span calibration
0–16777215
Other calibration commands
SCs.LC.CD
Set deadload coefficient
0–16777215
SCs.LC.CW
Set span coefficient
0–16777215
SCs.LC.CZ
Set temporary zero
0–16777215
SCs.REZERO
Rezero
0–16777215
SCs.LASTZERO
Last zero
0–16777215
SCs.WLIN.W1–SCs.WLIN.W5
Raw count value for linearization points 1–5
0–16777215
SCs.WLIN.V1–SCs.WLIN.V5
Test weight value for linearization points 1–5
0.000001–9999999
SCs.WLIN.C1–SCs.WLIN.C5
Calibrate linearization points 1–5
—
SCs.CALSPAN
Platform span calibration on
Cal-Match
scale types
Table 5-7.
iQUBE
2
Calibration Commands