Rice Lake iQUBE2 Digital Diagnostic Junction Box User Manual
Page 50

46
iQUBE
2
Installation Manual
See Section 6.0 on page 46 for more information about the commands listed in Table 5-11.
Command
n = digital I/O bit number
Description
Values
DIGIOn
Set digital I/O bit function
OFF, ZERO, TARE, NT/GRS, UNITS, PRINT, CLRCN,
CLRTAR, SETPNT, CELHLTH, HOSTCTL
DIGIOn.SCALE
Assign digital I/O bit n to scale
1–5 (scale number)
DIN#0
Read all digital inputs
Returns value 0–F (status of bits 1–4 represented as
hexadecimal values 1, 2, 4, 8)
DIN#n
Read digital input n
1=low, 0=high
DON#0
Set all digital outputs on (active)
—
DON#n
Set digital output n on
1–4
DOFF#0
Set all digital outputs off (inactive)
—
DOFF#n
Set digital output n off
1–4
MSDOUT
Reads all outputs in the system
—
Table 5-10.
iQUBE
2
Digital I/O Commands
Command
s = scale number
Description
Values
DIA.PS
Query power supply error
—
DIA.MSCONNECT
Query secondary connection error
—
DIA.PSEXC
Query excitation error
—
DIA.CELLCONNECT
Query cell connection error
—
DIA.FLAGS
Show all diagnostic flags simultaneously
See Section 6.2.4 on page 50
DIA.CLEAR
Clear all diagnostic flags
—
DIA.OVERLOAD
Query cell overload error
—
SCs.DIA.OVERLOAD
Set cell overload value
–100.00 – 100.00 (percent)
DIA.UNDERLOAD
Query cell underload error
—
SCs.DIA.UNDERLOAD
Set cell underload value
–100.00 – 100.00 (percent)
DIA.ZREF
Query zero reference error
SCs.DIA.ZREF
Enable zero reference diagnostic
OFF, ON
SCs.DIA.ZREF.RANGE
Set zero reference range
0.00 – 100.00 (percent)
SCs.DIA.ZREF.THRESH
Set zero reference threshold
–100.00 – 100.00 (percent)
SCs.DIA.ZREF.TIME
Set zero reference timer
1 – 60 (seconds)
DIA.DRIFT
Query cell drift error
—
SCs.DIA.DRIFT
Enable cell drift diagnostic
OFF, ON
SCs.DIA.DRIFT.RANGE
Set cell drift range
0.00 – 100.00 (percent)
SCs.DIA.DRIFT.THRESH
Set cell drift threshold
0.00 – 100.00 (percent)
SCs.DIA.DRIFT.TIME
Set cell drift timer
1 – 300 (seconds)
DIA.SDALL
Returns real-time load cell noise (in counts)
DIA.NOISE
Enable peak-to-peak noise diagnostic
OFF, ON
DIA.NOISE.THRESH
Set peak-to-peak noise threshold for all scales
0.01 – 1000 (microvolts)
DIA.NOISE.TIME
Set peak-to-peak noise timer
1 – 300 (seconds)
DIA.UNBAL
Query unbalanced load error
—
SCs.DIA.UNBAL
Enable unbalanced load diagnostic
OFF, ON
SCs.DIA.UNBAL.RANGE
Set unbalanced load range
5 – 75 (percent)
SCs.DIA.UNBAL.THRESH
Set unbalanced load threshold
0 – 50 (percent)
Table 5-11.
iQUBE
2
Diagnostic Commands