1 front panel calibration, Front panel calibration – Rice Lake iQUBE2 Digital Diagnostic Junction Box User Manual
Page 38

34
iQUBE
2
Installation Manual
4.1.1 Front Panel Calibration
From the 920i front panel, a theoretical, 2-point, or Cal-Match calibration can be performed. Calibration is per-
formed in real-time with each button press.
Theoretical Calibration
To perform a theoretical calibration, ensure the scale is empty and navigate to the
Theoretical Calibration
parameter
(see Figure 4-1). Then, press the
Calibrate
softkey. The dead load value is captured and the cells trimmed based on
the capacity and sensitivity values entered for the cells. Scale capacity is set at the total of the load cell capacities.
2-Point Calibration
2-point calibration is used for traditional scale calibration where load cells will not be trimmed.
To perform a 2-point calibration,
1. Navigate to the
CALIB
menu, then select
2-POINT
.
2. Press
ENTER
or
DOWN
( ) to go to the
WZero
submenu.
3. Ensure there is no weight on the scale, then press the
Calibrate
softkey to calibrate WZero. Or, press one of
the softkeys described below.
4. Press
ENTER
to accept the value and move to the next prompt (
WVAL
).
5. With
WVAL
displayed, press
ENTER
or
DOWN
( ) to view the stored test weight value. Use the numeric key-
pad to enter the desired value.
6. Press
ENTER
to advance to
WSPAN.
7. With
WSPAN
displayed, press
ENTER
or
DOWN
to view the A/D count. Place test weights on the scale equal
to WVAL. Press the
CALIBRATION
softkey to calibrate WSPAN.
8. Press
ENTER
to accept the value. Basic calibration is complete.
9. Press the
Save and Exit
softkey to return to weigh mode. To continue with linearization, see Section 4.1.2.
Cal-Match
®
Calibration (Corner, Section)
Cal-Match calibration can be performed using a vehicle as the test weight, but a smaller footprint and heavier
weight value will often yield better results. Mechanical characteristics of the scale may prevent Cal-Match from
automatically computing load cell trim factors. To perform a Cal-Match calibration,
1. Navigate to the
CALIB
submenu. Select which calibration type to use:
CORNER
or
SECTION
.
2. Press
ENTER
or
UP
( ) to go to the
WVAL
menu.
3. Press
ENTER
or
DOWN
( ) to view the stored test weight value. Use the numeric keypad to enter the desired
value. Press
ENTER
to accept the value and advance to Cal-Match.
CALIB Menu
Level 4 Submenus
Menu
Description
THEORET
Provides access to perform a theoretical calibration, which mathematically calibrates the scale using load
cell sensitivities, and setting full scale capacity based on the sum of the load cell capacities.
WZERO
Provides access to perform a traditional zero calibration for a 2-Point scale. Advanced options are also
available. See Section 4.1.2 on page 35)
WVAL
Provides access to display and edit the test weight value.
WSPAN
Provides access to display, calibrate and edit the traditional span calibration value for a 2-Point scale.
CALZERO
Advanced options for specifying an alternative zero source for a
Cal-Match
®
scale.
CALSPAN
Provides access to perform an optional, final span calibration on a trimmed scale.
WLIN
POINT 1 -
POINT 5
Provides access to test weight and calibration values for up to five linearization Provides access to test
weight and calibration values for up to five linearization points. Perform linear calibration only after Zero and
Span calibration points have been set. See Section 4.1.2 on page 35 for more information about
linearization.
REZERO
Provides access to perform an optional, final zero calibration, which removes an offset value from the zero
and span calibration values.
Table 4-2. CALIB menu parameters