0 diagnostics, 1 diagnostic functions, 1 diagnostic error detection – Rice Lake iQUBE2 Digital Diagnostic Junction Box User Manual
Page 51: Diagnostics, Diagnostic error detection

Diagnostics
47
6.0 Diagnostics
The advanced diagnostic capabilities of the iQUBE
2
include power supply and excitation voltage monitoring, load
cell connection faults, cell overload or underload, unbalanced load, load cell drift, load cell peak-to-peak noise and
zero reference. Diagnostic flexibility allows for simple configuration and the ability to include the diagnostics in
the weight data or to monitor diagnostics on a separate communication port from weight data.
Load Cell Status
The iQUBE
2
CPU board provides bi-color cell status LEDs next to each load cell connector. These LEDs indicate
the status of connected load cells.
6.1
Diagnostic Functions
The iQUBE
2
provides three types of diagnostic functions: system, cell, and scale diagnostics.
• System diagnostics monitor the power supply and excitation voltages and communications between primary
and secondary boards. System diagnostics are always enabled.
• Cell diagnostics monitor cell connection, cell overload, and cell underload conditions. Cell diagnostics are
enabled for all cells that are assigned to a scale.
• Scale diagnostics provide control parameters to allow configuration for specific scale applications and must be
explicitly enabled. Scale diagnostic commands provide monitoring for zero reference, cell drift, peak-to-peak
noise, and unbalanced load conditions.
6.1.1 Diagnostic Error Detection
Diagnostics generate a code that can be included in any of the six available data formats, FMT1 through FMT6 (see
Section 7.0 on page 52). By default, the diagnostic code token,
ure in a scale system may generate several error conditions, only the most critical diagnostic code is shown at one
time. Table 6-2 shows the diagnostic codes for the various error types, in priority order.
LED Color
Meaning
Green
Load cell good
Red
Load cell error
Red/green flashing
Consistent error being cleared by host
Off
Disabled/load cell not assigned to a scale
Table 6-1. LED Status Indications
Diagnostic
Code
Error Type
Associated Query
Command
P
Power supply
DIA.PS
S
Secondary communication
DIA.MSCONNECT
E
Excitation
DIA.PSEXC
C
Cell connection
DIA.CELLCONNECT
R
Zero reference
DIA.ZREF
V
Overloaded cell
DIA.OVERLOAD
D
Drift
DIA.DRIFT
U
Underloaded cell
DIA.UNDERLOAD
N
Peak noise
DIA.NOISE
L
Unbalanced cell
DIA.UNBAL
No error
—
Table 6-2. Diagnostic Codes and Query Commands