5.1.5
Normal Mode Commands
The serial transmit weight data commands (see Table 5-14) transmit data to the EDP port on demand. The SX
and EX commands are valid only in normal operating mode; all other commands are valid in either setup or
normal mode.
5.1.6
Batching Control Commands
The commands listed below provide batching control through the EDP port.
BATSTART
If the BATRUN digital input is active (low) or not assigned, The BATSTART command can be used to start the
batch program.
BATRESET
Stops the program and resets the batch program to the first batch step. Run the BATRESET command after
making changes to the batch configuration.
BATPAUSE
Stops the batch program at the current step. All digital outputs set on by the current step are set off. The BATSTA
DIGIN, BATSTART EDP command can be used to restart the batch program at the current step.
BATSTATUS
The BATSTATUS command is used to check the current status of various setpoint and batching conditions.
BATSTATUS returns 14 bytes of status data as described in Figure 5-15 on page 51. BATSTATUS is principally
used to provide status information to a controlling batch program when using the Remote I/O Interface option.
Status information returned in bytes 3 - 12 is coded as ASCII characters @ (hex 40) through 0 (hex 4F); only the
low order bits of these characters are significant. Table 5-15 on page 51 shows the low order bit assignments for
bytes 3 - 12. Use Table 5-16 on page 51 to interpret the status of the low order bits for a given ASCII character.
AO2.TWSPAN
Span calibration
0–65535
Command
Description
Response Format
SX
Start EDP streaming
OK or ??
EX
Stop EDP streaming
OK or ??
RS
Reset system
—
XG
Transmit gross weight in displayed units
nnnnnn UU
where nnnnnn is the weight value, UU is the units.
XN
Transmit net weight in displayed units
XT
Transmit tare weight in displayed units
XG2
Transmit gross weight in non-displayed units
XN2
Transmit net weight in non-displayed units
XT2
Transmit tare weight in non-displayed units
XE
Query system error conditions
xxxxx yyyyy
See Section 8.1.2 on page 61 for detailed information about
the XE command response format.
XA
Transmit accumulation value
UID
set unit ID
nnnnnn
CONSNUM
Set consecutive number
Table 5-14. Normal Mode EDP Commands