Gh 5-13 for more information – Rice Lake 320IS Plus Intrinsically Safe Digital Weight Indicator - Installation Manual User Manual
Page 55
320IS Plus Installation Manual - EDP Commands
49
DIGOUT
Digital output
NONE, 1-4
RELNUM
Relative setpoint number
1-8
START
Starting setpoint
1-8
END
Ending setpoint
1-8
BATCHNG
Batching mode
OFF, AUTO, MANUAL
Command
Description
Values
DIGIN1
DIGIN2
DIGIN3
DIGIN4
Digital input function
OFF, ZERO, TARE, NT/GRS, UNITS, DSPTAR, PRINT, CLRCN,
KBDLOC, HOLD, BATSTA, BATPAU, CLRTAR, CLRACC, ACCUM,
ENTER, BATRUN, GROSS, NET, PRIM, SEC, NEWID
Table 5-10. DIG IN EDP Commands
EDP digital output commands only function in TEST mode (see Figure 8-4 on page 70).
Command
Description
DON#n
Set digital output n on
DOFF#n
Set digital output n off
DOFF#0
Set all digital outputs off
DON#0
Set all digital outputs on
For commands ending with “#n”, n is the digital output (1–4) being set on or off.
Table 5-11. DIG OUT EDP Commands
Command
Description
Values
AO1.SOURCE
Analog output source
GROSS, NET
AO1.OFFSET
Zero offset
0%, 20%
AO1.ERRACT
Error action
FULLSC, HOLD, ZEROSC
AO1.MIN
Minimum value tracked
0–999 990
AO1.MAX
Maximum value tracked
0–999 990
AO1.TWZERO
Zero calibration
0–65535
AO1.TWSPAN
Span calibration
0–65535
Table 5-12. ALGOUT 1 EDP Commands
Command
Description
Values
AO2.SOURCE
Analog output source
GROSS, NET
AO2.OFFSET
Zero offset
0%, 20%
AO2.ERRACT
Error action
FULLSC, HOLD, ZEROSC
AO2.MIN
Minimum value tracked
0–999 990
AO2.MAX
Maximum value tracked
0–999 990
AO2.TWZERO
Zero calibration
0–65535
Table 5-13. ALGOUT 2 EDP Commands
Command
Description
Values
Table 5-9. SETPNT EDP Commands