3 the resetconfiguration command, 4 parameter setting commands, The resetconfiguration command – Rice Lake 320IS Plus Intrinsically Safe Digital Weight Indicator - Installation Manual User Manual
Page 52: Parameter setting commands
46
320IS Plus Installation Manual
5.1.3
The RESETCONFIGURATION Command
The RESETCONFIGURATION command can be used to restore all configuration parameters to their default
values. Before issuing this command, the indicator must be placed in test mode (press and hold setup switch until
TEST
is displayed).
This command is equivalent to using the DEFLT function on the TEST menu. See Section 8.8 on page 70 for
more information about test mode and using the TEST menu.
All load cell calibration settings are lost when the RESETCONFIGURATION command is run.
5.1.4
Parameter Setting Commands
Parameter setting commands allow you to display or change the current value for a particular configuration
parameter See Tables 5-3 through 5-13 for more information on menu parameters.
Current configuration parameter settings can be displayed in either setup mode or normal mode using the
following syntax:
command
Most parameter values can be changed in setup mode only. Use the following command syntax when changing
parameter values:
command=value
where
value
is a number or a parameter value. Use no spaces before or after the equal (=) sign. If you type an
incorrect command or value, the display reads
??
. Changes to the parameters are saved as they are entered but
typically do not take effect until you exit setup mode.
For example, to set the motion band parameter to 5, type the following:
MOTBAND=5D
ZZ
Display current weight and annunciator
status. See Section 8.2.2 on page 63 for
more information.
S
Display one frame of stream format
Command
Description
Values
GRADS
Graduations
1–100 000
ZTRKBND
Zero track band
OFF, 0.5D, 1D, 3D
ZRANGE
Zero range
1.9%, 100%
MOTBAND
Motion band
1D, 2D, 3D, 5D, 10D, 20D, OFF
OVRLOAD
Overload
FS+2%, FS+1D, FS+9D, FS
DIGFLTRX*
Digital filtering
1, 2, 4, 8, 16, 32, 64
DFSENS
Digital filter cutout sensitivity
2OUT, 4OUT, 8OUT, 16OUT, 32OUT, 64OUT, 128OUT
DFTHRH
Digital filter cutout threshold
NONE, 2DD, 5DD, 10DD, 20DD, 50DD, 100DD, 200DD, 250DD
TAREFN
Tare function
BOTH, NOTARE, PBTARE, KEYED
SMPRAT
Sample rate
15Hz, 30Hz, 60Hz, 7.5Hz
* X = 1, 2, 3
Table 5-3. CONFIG EDP Commands
Command
Function
Table 5-2. EDP Reporting Commands
Note