beautypg.com

Procedure – LaMotte SMART Spectro Spectrophotometer User Manual

Page 98

background image

PROCEDURE

Use 10 mm square cell adapter

1. Press and hold ON button until spectrophotometer turns on.
2. Scroll to and select PROGRAMMED TESTS.
3. Scroll to and select ALL TESTS (or another sequence containing

16 Cl Free-UDV) from TESTING MENU.

4. Scroll to and select 16 Cl Free-UDV from menu.
5. Rinse a clean vial (0156) with sample water.
6. Use the syringe (1184) to add 3mL of sample to the vial.
7. Insert the vial into chamber, close the lid and select SCAN BLANK.
8. Remove the vial from the Spectro.
9. Use the syringe (1184) to add 3mL of sample to a *Free Chlorine UDV

(4311).

10. Shake vigorously until powder dissolves completely.

NOTE: If powder residue remains in the bottom of the vial after inverting or air

bubbles form, invert once more and tap bottom of vial sharply once or twice to

dislodge powder and bubbles. Mix.

11. Immediately insert tube into chamber, close lid and select SCAN SAMPLE.

Record result in ppm free chlorine.

12. Press OFF button to turn the spectrophotometer off or press EXIT button to

exit to a previous menu or make another menu selection.

NOTES: For best possible results, a reagent blank should be determined to

account for any contribution to the test result by the reagent system. To determine

the reagent blank, follow the above test procedure to scan a distilled or deionized

water blank. Then follow the above procedure to perform the test on a distilled or

deionized water sample. This test result is the reagent blank. Subtract the reagent

blank from all subsequent test results of unknown samples. It is necessary to

determine the reagent blank only when a new lot number of reagents is obtained.

UDVs from opened pouches should be used promptly. Store unused vials from

opened pouches in the Foil Storage Bag (9467) to extend the shelf life of the

reagent. Generally, UDVs stored in the bag should be used within 10 days if the

humidity is less than 50% and within 5 days if humidity is greater than 50%. The Foil

Storage Bag contains a desiccant pack with indicator. When the indicator in the

window turns from blue to pink, the bag should be replaced.

SMART Spectro Test Procedures 2.11

CHLORINE-FREE, UDV

Test P

rocedures