LaMotte SMART Spectro Spectrophotometer User Manual

Page 45

background image

SMART Spectro Operator’s Manual 2.11

45

4. Press

*ENTER to select SELECT

WL

SELECT WL

WL=560 NM

*

5. Press or

to scroll to desired

wavelength.

*

SELECT WL

WL=600 NM

6. Press

*ENTER to select the

wavelength. The meter is ready to

scan.

%T/ABS 600 NM
SELECT WL

*

SCAN BLANK
SCAN SAMPLE
END OF LIST

7. Insert the blank into the chamber.

Press

*ENTER to scan the blank.

Note: For most %T/ABS tests, a

clear colorless water blank should

be used.

%T/ABS 600 NM
SELECT WL
SCAN BLANK

*

SCAN SAMPLE
END OF LIST

8. Insert a reacted sample into the

chamber. Press

*ENTER to scan

the sample.

%T/ABS 600 NM
T=90.7%T
A=0.0424A
PRINT PRESS ENTER

At this point, it is possible to scan another sample, scan another blank, or select

another wavelength. To print the results to a connected computer or serial

printer, press

*ENTER and return to previous menu. Press EXIT to escape to

previous menu.
If no more samples are to be scanned, press

EXIT to return to the MAIN MENU

or press

OFF to turn off the spectrophotometer.