PROCEDURE

Use universal sample holder.

1. Press and hold ON button until spectrophotometer turns on.
2. Scroll to and select PROGRAMMED TESTS from menu.
3. Scroll to and select ALL TESTS (or another sequence containing

20 CHLOR DIOX) from TESTING MENU.

4. Scroll to and select 20 CHLOR DIOX from menu.
5. Rinse a clean tube (0290) with sample water. Fill to 10 mL line with sample.
6. Insert tube into chamber, close lid and select SCAN BLANK.
7. Remove tube from Spectro. Pour out all but a suffi cient amount of sample

water to cover tablet. Add 5 drops of Glycine Solution (6811).

8. Add one *Chlorine DPD #1 Instrument Grade Tablet (6903A). Crush tablet

with tablet crusher. Cap and shake until tablet dissolves. Fill to 10 mL line

with sample water. Solution will turn pink if chlorine dioxide is present. Wait 15

seconds, but no longer than 30 seconds. Mix.

9. Insert tube into chamber, close lid and select SCAN SAMPLE.
10. Press OFF button to turn spectrophotometer off or press EXIT button to exit to

a previous menu or make another menu selection.

CHLORINE DIOXIDE

SMART Spectro Test Procedures 2.11

Test P

rocedures