LaMotte SMART Spectro Spectrophotometer User Manual
Page 41

SMART Spectro Operator’s Manual 2.11
41
6. Place a blank in the sample
chamber. Press
*ENTER to SCAN
BLANK. (STD. 1 is always a blank.)
The display will indicate that the
blank is being scanned. The
display will indicate BLANKED. A
cursor will appear on the character
selection line.
ENTER STD. 1
*
SCAN BLANK
STD.1 =
!
0 1 2 3 4 5 6 7 8 9 .
7. Place the reacted standard of
the lowest concentration in the
chamber. Press
or
to enter
the concentration of the standard
by selecting the characters with
the cursor.
ENTER STD. 1
*
SCAN BLANK
STD.1 =
!
0
1 2 3 4 5 6 7 8 9
8. After the concentration has been
entered, scroll to !. Press
*ENTER.
The standard will be scanned
and the absorbance will be
displayed.
ENTER STD. 1
*
BLANKED
STD.1 = 1.0
A=0.0016A
9. Press
EXIT to return to the New
Calibration menu.
NEW CALIBRATION
*
ENTER STD. 1
ENTER STD. 2
ENTER STD. 3
ENTER STD. 4
10. Press
ENTER and use the same
procedure to scan the second
data point. Repeat the procedure
until the data from all of the
standards has been entered into
the calibration.
Note: Input the number of data
points previously selected.
For example, when inputting 6
standards ignore the request for
ENTER STD. 7.
NEW CALIBRATION
ENTER STD. 4
ENTER STD. 5
ENTER STD. 6
*
ENTER STD. 7