SMART Spectro Operator’s Manual 2.11

25

will then move the grating a precise predetermined amount from this position.

This precise movement will cause the grating to be positioned at 546 nm every

time. Once calibrated the wavelength displayed during testing is accurate to ±2

nm. When the wavelength calibration is complete the display will go back to the

Main Menu.

PROGRAMMED TESTS

INTRODUCTION

The PROGRAMMED TESTS mode is used to run all LaMotte pre-programmed

tests and USER TESTS. This is also where USER TESTS and SEQUENCES are

set-up and edited.

1. Press

ON. The LaMotte logo

screen will appear for about 2

seconds and the MAIN MENU will

appear.

MAIN MENU 12:00:01

*

CALIBRATE WL
PROGRAMMED TESTS
%T/ABS
PC LINK

2. Press or

to scroll to

PROGRAMMED TESTS.

MAIN MENU 12:00:01
CALIBRATE WL

*

PROGRAMMED TESTS
%T/ABS
PC LINK

3. Press

*ENTER to select

PROGRAMMED TESTS. In

the

PROGRAMMED TESTS

menu there are three alterable

sequences and one

ALL TESTS

fi xed sequence as well as the

EDIT

function.

MAIN MENU 12:00:01

*

CALIBRATE WL
PROGRAMMED TESTS
%T/ABS
PC LINK