Calibration – LaMotte SMART Spectro Spectrophotometer User Manual

Page 24

background image

24

SMART Spectro Operator’s Manual 2.11

inner menu to the

MAIN MENU by repeatedly pushing the EXIT button. Pushing

OFF at any time will turn the spectrophotometer off.

CALIBRATION

CALIBRATE WAVELENGTH

The Calibrate Wavelength (CALIBRATE WL) mode is used to establish or

re-establish the accuracy of the wavelength selection process. Normally, the

Calibrate Wavelength procedure should be run after the SMART Spectro is

turned ON and allowed to warm up for 15 minutes or if operating conditions

(temperature, humidity, etc.) change signifi cantly.
For fi eld use, when operating with the battery, calibrate wavelength prior to

going into the fi eld using AC power. This will increase battery life in the fi eld.

Alternatively calibrate wavelength in the fi eld immediately before testing. Turn

Spectro on immediately before scanning blank. Calibrate wavelength just before

scanning blank.

1. Press

ON. The LaMotte logo

screen will appear for about 2

seconds and the MAIN MENU will

appear.

MAIN MENU 12:00:01

*

CALIBRATE WL
PROGRAMMED TESTS
%T/ABS
PC LINK

2. Press or

to scroll to

CALIBRATE WL.

MAIN MENU 12:00:01

*

CALIBRATE WL
PROGRAMMED TESTS
%T/ABS
PC LINK

3. Press

*ENTER to select

CALIBRATE WL.

MAIN MENU 12:00:01

*

CALIBRATE WL
PROGRAMMED TESTS
%T/ABS
PC LINK

The Calibrate Wavelength procedure takes about 1-2 minutes to be completed.

During the calibration, the Spectro will display two numbers at the bottom of the

screen. The fi rst number is fi xed. The second number will change and can have

a range of values. The microprocessor will move the grating in search of the

position that gives a very specifi c maximum light intensity. The microprocessor