Procedure – LaMotte SMART Spectro Spectrophotometer User Manual
Page 142
PROCEDURE
Use 10 mm square cell adapter.
1. Press and hold ON button until spectrophotometer turns on.
2. Scroll to and select PROGRAMMED TESTS.
3. Scroll to and select ALL TESTS (or another sequence containing
33 Copper-UDV) from TESTING MENU.
4. Scroll to and select 33 Copper-UDV from menu.
5. Rinse a clean vial (0156) with sample water.
6. Use the syringe (1184) to add 3 mL of sample to the vial.
7. Insert the vial into chamber, close lid and select SCAN BLANK.
8. Remove vial from Spectro.
9. Use the syringe (1184) to add 3 mL of sample to a Copper UDV vial
(4314).
10. Wait 2 minutes.
11. Invert vial 3 times to mix.
NOTE: If powder residue remains in the bottom of the vial after inverting, or air
bubbles form, invert once more and tap bottom of vial sharply once or twice to
dislodge powder or bubbles. Mix.
12. Insert tube into chamber, close lid and select SCAN SAMPLE. Record
result.
13. Press OFF button to turn the spectrophotometer off or press EXIT button to
exit to a previous menu or make another menu selection.
NOTE: For best possible results, a reagent blank should be determined to account
for any contribution to the test result by the reagent system. To determine the
reagent blank, follow the above test procedure to scan a distilled or deionized
water blank. Then follow the above procedure to perform the test on a distilled or
deionized water sample. This test result is the reagent blank. Subtract the reagent
blank from all subsequent test results of unknown samples. It is necessary to
determine the reagent blank only when a new lot number of reagents is obtained.
UDVs from opened pouches should be used promptly. Store unused vials from
opened pouches in the Foil Storage Bag (9467) to extend the shelf life of the
reagent. Generally, UDVs stored in the bag should be used within 10 days if the
humidity is less than 50% and within 5 days if humidity is greater than 50%. The
Foil Storage Bag contains a dessicant pack with indicator. When the indicator in the
window turns from blue to pink, the bag should be replaced.
SMART Spectro Test Procedures 2.11
COPPER, UDV
Test P
rocedures