APPLICATION:

Domestic and industrial wastes.

RANGE:

0–15,000 mg/L COD

MDL:

500 mg/L

METHOD:

Dichromate in the presence of silver salts, at high

temperature in a closed system, oxidizes most

organic compounds to 95-100% of the theoretical

amount. This process is called digestion. As

dichromate oxidizes the organic compounds, a

green complex is formed. The concentration of the

green complex is measured at 605 nm and is directly

proportional to the COD of the sample.

SAMPLE HANDLING &

PRESERVATION:

Collect samples in glass and test as soon as

possible. If samples must be stored, preservation is

accomplished by the addition of concentrated H

2

SO

4

to adjust the pH below 2. Samples with suspended

solids should be homogenized in a blender (100

mL for 30 seconds) and then stirred gently with a

magnetic stirrer.

INTERFERENCES:

Volatile organic compounds are not oxidized to the

extent that they are in the vapor above the digestion

solution.

Therefore, they do not contribute to the COD reading.

Contains mercury sulfate to prevent interference

from chloride. Nitrite gives a positive interference of

1.1 ppm O

2

per ppm NO

2

–N, which is insignifi cant

unless nitrite concentrations are very high. Other

reduced inorganic compounds are stoichiometrically

oxidized, causing a positive interference. Corrections

can be made for these compounds based upon their

stoichiometry and concentrations.