SMART Spectro Operator’s Manual 2.11

ENTERING A NEW CALIBRATION

1. Press to scroll to

NEW

CALIBRATION.

102 H2O

NAME THE TEST
SELECT WL

*

NEW CALIBRATION
FORMAT RESULT

2. Press

*ENTER to select NEW

CALIBRATION.

INPUT DATA NUMBERS

DATA NUMBERS = 5

SET PRESS ENTER

3. Press or

to select the

number of data points that will

be used to create the calibration

curve. The maximum number of

data points is 8.

INPUT DATA NUMBERS

DATA NUMBERS = 6

SET PRESS ENTER

4. Press

*ENTER to save the

selection.

NEW CALIBRATION

*

ENTER STD.1
ENTER STD.2
ENTER STD.3
ENTER STD.4

5. Press

*ENTER to select the fi rst

data point.

ENTER STD. 1

*

SCAN BLANK
STD.1 =

!0123456789.