LaMotte SALT WATERPROOF POCKETESTER User Manual
Salt waterproof pocketester for pools
SALT WATERPROOF POCKETESTER
FOR POOLS
CODE 5-0078
QUANTITY
CONTENTS
CODE
120 mL
Calibration Standard, 3.0 ppt
6005-J
1
Salt PockeTester
26094-38
To order individual reagents or test kit components, use the specified code
number. NOTE: If a 500 mL container of the 3.0 ppt Calibration Standard
is needed order Code 6005-L.
PRECONDITIONING
Remove the electrode cap. Soak the electrodes for a few minutes in alcohol
to remove oils.
CALIBRATION
The Salt Tester should be calibrated before use. The tester should be
calibrated on a regular basis. Once a week is recommended.
NOTE:
The Salt PockeTester measures in part per thousand ppt which is
1 part per thousand (ppt) = 1000 parts per million (ppm)
Example: 3.1 ppt = 3,100 ppm
CALIBRATION PROCEDURE
1.
Open the battery compartment lid (end with the lanyard loop). The
two white buttons are Increment (INC) and Decrement (DEC)
Calibration keys.
2.
Rinse the electrode in deionized water and then rinse it in the
Calibration Standard (6004).
3.
Place the electrode in a small container of the Calibration Standard.
The cap of the Tester may be used for this. Fill the cap half full. Make
sure the sensor is fully immersed.
4.
Press ON/OFF to turn the meter on. Wait several minutes for the
display to stabilize.
5.
Press the INC or DEC buttons to adjust the reading to match the
Calibration Standard value (3.0 ppt).
6.
After 3 seconds without a button press the display will flash three
times and then show “ENT”, indicating that the meter has accepted
the calibration standard. The meter will then automatically return to
the measurement mode.
7.
Replace the battery cap.