COLOR

PLATINUM COBALT METHOD

NO REAGENTS REQUIRED

Color in water may be attributed to humus, peat, plankton, vegetation, and natural

metallic ions, such as iron and manganese, or industrial waste. Color is removed

to make water suitable for domestic and industrial use. Color may have to be

removed from industrial waste before it is discharged to a waterway.

APPLICATION:

Potable water and water with color due to natural

materials.

RANGE:

0–1,000 color units

MDL:

15 cu

METHOD:

Color is determined by a meter that has been calibrated

with colored standards of known platinum cobalt

concentration. True color, the color of water in which the

turbidity has been removed, is measured.

SAMPLE HANDLING &

PRESERVATION:

Collect all samples in clean glassware. Determine color

as soon as possible to avoid biological or chemical

changes that could occur in the sample during storage.

INTERFERENCES:

Turbidity will interfere. Filter before testing.

SMART Spectro Test Procedures 2.11

COLOR

Test P