Procedure – LaMotte SMART Spectro Spectrophotometer User Manual
Page 137

PROCEDURE
Use universal sample holder.
1. Press and hold ON button until spectrophotometer turns on.
2. Scroll to and select PROGRAMMED TESTS.
3. Scroll to and select ALL TESTS (or another sequence containing
31 Cu-Cuprizone) from TESTING MENU.
4. Scroll to and select 31 Cu-Cuprizone from menu.
5. Rinse a tube (0290) with sample water. Fill to the 10 mL line with sample.
6. Insert the tube into chamber, close lid and select SCAN BLANK.
7. Remove tube from Spectro and add 5 drops of Copper A (P-6367). Cap and
mix.
8. Add 5 drops of *Copper B (P-6368). Cap and mix.
9. Wait 5 minutes. Mix.
10. Insert tube into chamber, close lid and select SCAN SAMPLE. Record
result.
11. Press OFF button to turn the spectrophotometer off or press EXIT button to
exit to a previous menu or make another menu selection.
NOTES: For best possible results, a reagent blank should be determined to
account for any contribution to the test result by the reagent system. To determine
the reagent blank, follow the above test procedure to scan a distilled or deionized
water blank. Then follow the above procedure to perform the test on a distilled or
deionized water sample. This test result is the reagent blank. Subtract the reagent
blank from all subsequent test results of unknown samples. It is necessary to
determine the reagent blank only when a new lot number of reagents is obtained.
The reaction may stain the tubes. Scrub tubes thoroughly after each use.
COPPER, Cuprizone
SMART Spectro Test Procedures 2.11