beautypg.com

Phosphate–high range – LaMotte SMART Spectro Spectrophotometer User Manual

Page 224

background image

PHOSPHATE–HIGH RANGE

VANADOMOLYBDOPHOSPHORIC ACID METHOD

CODE 3655-SC

QUANTITY

CONTENTS

CODE

4 x 30 mL

* VM Phosphate Reagent

*4410-G

1

Pipet, 1.0 mL, plastic

0354

*WARNING: Reagents marked with an * are considered to be potential health

hazards. To view or print a Material Safety Data Sheet (MSDS) for these reagents

go to www.lamotte.com. To obtain a printed copy, contact LaMotte by e-mail,

phone or fax.

Phosphate treatments in boiler and cooling water and other industrial water

systems are run at levels up to 100 ppm orthophosphate. These high levels permit

the use of a simpler, high range test.

APPLICATION:

Boiler, cooling, and industrial waters.

RANGE:

0.0–70.0 ppm Phosphate

MDL:

1.0 ppm

METHOD:

Orthophosphate reacts in acid conditions

with ammonium vanadomolybdate to form

vanadomolybdophosphoric acid. This yellow color is

proportional to the concentration of orthophosphate and

is read colorimetrically.

SAMPLE HANDLING

& PRESERVATION:

If the analysis cannot be performed the same day of

collection, the sample should be preserved by the

addition of 2 mL of concentrated sulfuric acid or 40 mg

mercuric chloride per liter and refrigerated at 4°C.

INTERFERENCES:

Silica interferes only if the sample is heated. Arsenate,

fl uoride, thorium, bismuth, sulfi de, thiosulfate, and

thiocyanate cause negative interference.

SMART Spectro Test Procedures 2.11

PHOSPHATE, High Range

Test P

rocedures