Test mode – LaMotte SMART Spectro Spectrophotometer User Manual
Page 50
50
SMART Spectro Operator’s Manual 2.11
5. Press
*ENTER to select the store
method and return to the
MAIN
MENU.
MAIN MENU 12:00:01
*
CALIBRATE WL
PROGRAMMED TESTS
%T/ABS
PC LINK
TEST MODE
Select
TEST MODE from the Main Menu. Test Mode allows the user to change
the way that blanking and reading of results is carried out during the
%T/ABS
testing. It does not affect any of the Pre-programmed tests.
REGULAR is the
default test mode and should be used whenever the blank is clear or has less
absorbance than the samples. This will be the case most of the time. In the
rare instances where the blank will be darker or have more absorbance than the
samples the Test Mode must be switched to
REVERSE to avoid error messages
and incorrect readings. Always reset the test mode to
REGULAR at the end of
each testing session.
The Test Mode should always be reset to REGULAR at the end of each testing
session.
1. Press
ON. The LaMotte logo
screen will appear for about 2
seconds and the MAIN MENU will
appear.
MAIN MENU 12:00:01
*
CALIBRATE WL
PROGRAMMED TESTS
%T/ABS
PC LINK
2. Press or
to scroll to
TEST MODE.
MAIN MENU 12:00:01
EDIT CLOCK
ENERGY MODE
STORE METHOD
*
TEST MODE
3. Press
*ENTER to select TEST
MODE.
TEST MODE
*
REGULAR
REVERSE
END OF LIST