beautypg.com

LaMotte pHTestr Spear User Manual

LaMotte Equipment

background image

INSTRUCTION MANUAL

pHTestr 10, 20, 30, 10BNC, Spear

Large Screen

Waterproof pH / Temperature Tester

Double Junction

Introduction

Thank you for selecting our microprocessor waterproof pH tester with USA or NIST

buffer set selection. You have one of five models:

• pHTestr10

• pHTestr20

• pHTestr30

• pHTestr10BNC

• pHSpear

This manual provides a step-by-step guide to operate the testers.

Before you begin:

Condition your pHTestr 10, 20, 30 electrodes by immersing it in electrode storage

solution or tap water for at least 30 minutes before use. DO NOT use de-ionized water.

Ensure that your pHSpear electrode is always soaked in the electrode storage solution

or tap water via its protective cap.

Note: For pHTestr10BNC, please refer to the pH electrode’s instruction manual.

pH Buffer Set Selection

Your tester features USA (pH 4.01, pH 7.00 and pH 10.01) or NIST (pH 4.01, pH 6.86,

and pH 9.18) standards. Select either one to suit your requirements.

1. While pressing the HOLD/ENT button, switch on the tester by pressing the ON/OFF

button.

2. Release the HOLD/ENT button. The display will flash either USA or NIST.

3. Press CAL button to toggle between the two buffer set standards.

4. Press the HOLD/ENT button to confirm the selection of the buffer set.

This Instruction Manual is also available for download on our Web-site: eutechinst.com or 4oakton.com

Figure 5: User Reset Sequence

With meter
powered off,
press CAL and
ON/OFF at the
same time.
First release
ON/OFF
button, then
CAL button.

After reset,
whole LCD
segment lits
up and goes
back to
measurement
mode.

Figure 4: Example of HOLD Function

pH Calibration

Calibration should be done regularly, preferably once a week. You can calibrate up to

three points using either the USA or the NIST buffer set standards.

1. Press ON/OFF button to switch unit on.

2. Dip electrode about 2 to 3 cm into the pH standard buffer solution.

3. Press the CAL button to enter calibration mode. The ‘CAL’ indicator will be

shown. The upper display will show the measured reading based on the last

calibration while the lower display will indicate the pH standard buffer solution.

Note: All testers have dual display during calibration mode

Note: To abort calibration, press the ‘CAL’ button.

4. Allow about 2 minutes for the tester reading to stabilize before pressing the

HOLD/ENT button to confirm the first calibration point. The upper display will

be calibrated to the pH standard buffer solution and the lower display will then

be toggling in between readings of the next pH standard buffer solution.

5. Repeat with other buffers if necessary. Rinse electrode in tap water before

dipping into next buffer.

Note: The calibration mode allows you to perform up to three calibration points

before returning to the measurement mode automatically. However, if you opted to

have only one or two calibration points, simply skip the remaining calibration points

by exiting to the measurement mode by pressing the CAL button.

pH Measurement

1

. Press the ON/OFF button to switch the tester on.

2. Dip the electrode about 2 to 3 cm into the test solution. Stir and let the

reading stabilize. For pHSpear, pierce the penetrating tip electrode through

your semi solid sample as per the desired depth. Rotate left and right several

times and tilt to ensure sample contact.

3. Note the pH value or press HOLD/ENT button to freeze the reading. To release

the reading, press HOLD/ENT again.

4. Press ON/OFF to turn off tester. If you do not press a button for 8.5 minutes,

the tester will automatically shut off to conserve batteries.

With meter
powered off,
press
HOLD/ENT
and ON/OFF
at the same
time. First release
ON/OFF button,
then HOLD/ENT
button.

After buffer
selection,
whole LCD
segment lits
up and goes
back to
measureme
nt mode.

Figure 1: Buffer Selection Sequence

To do a 1 point
calibration only, press
CAL button at this point
to exit to the
measurement mode.
Otherwise, proceed to
second buffer for a
second buffer for a
second point calibration.

To do a 2 point calibration only,
press CAL button at this point to
exit to the measurement mode.
Otherwise, proceed to third buffer
for a third buffer for a third point
calibration.

After the third point calibration, the
meter will automatically return to the
measurement mode. At any point, an
error message ‘Er. 1’ will be displayed
momentarily if the confirmed pH value
is not within the pH calibration window.

Figure 2: Example of pH Calibration Sequence

User Reset

You can reset the pH calibration to the factory default by using the user reset function.

Buffer set selection and temperature user calibration (pHTestr30) are not affected by the

user reset function.

1. Switch off the tester.

2. While pressing the ‘CAL’ button, press and release the ON/OFF button to enter the

‘User Reset’ selection menu. The screen will display ‘rSt’ on the bottom display with

a flashing ‘nO’ selection.

3. Use the ‘CAL’ button to toggle between ‘nO’ and ‘YES’ selection.

• nO deactivates reset selection

• YES activates the reset selection

4. Press the HOLD/ENT button to confirm the selection made.

5. If you have selected ‘YES’, the unit will show ‘CO’ momentarily and proceed to the

measurement mode with the calibration reset back to factory default value.

6. If ‘nO’ is selected, the unit will proceed to the measurement mode without any

calibration reset performed.

HOLD Function

This feature lets you freeze the display for a delayed observation

1. Press HOLD/ENT button to freeze the measurement. A ‘HOLD’ indicator will be

displayed and the measurement will be frozen.

2. Press HOLD/ENT again to release the measurement. The ‘HOLD’ indicator will not be

displayed anymore indicating the held measurement is released.

Automatic Temperature Compensation (ATC)

ATC feature is available only in pHTestr 10, pH Testr 20 and pH Testr 30. Through its in-

built temperature sensor, the measurement error due to the changes in electrode sensitivity

due to changes in temperature is compensated to give the actual pH reading of the

sample measured.

pH Testr10BNC and pH Spear has no ATC function and therefore the error should be

calculated to derive the actual pH reading of the sample. For these testers, the impact

on temperature compensation is 0.003 pH /°C / pH away from pH 7. For example, if

pHSpear is calibrated at room temperature (25 °C) and is measuring sample around pH 4

at around 5 °C.

Temperature difference : 25°C - 5°C = 20 °C

pH away from neutral : 7pH - 4pH = 3 pH

Total error

: 0.003 x 20 x 3 = 0.18 pH

This error value of 0.18 pH should be taken into consideration to derive the actual pH

reading of the sample.

This manual is related to the following products: