COBALT
PAN METHOD • CODE 4851
QUANTITY
CONTENTS
CODE
60 mL
*Cobalt Buffer
*4852-H
60 mL
*Cobalt Indicator Reagent
*4853-H
30 mL
*Stabilizer Solution
*4854-G
2
Pipet, 1.0 mL, plastic
0354
1
Pipet, 0.5 mL, plastic
0353
*WARNING: Reagents marked with an * are considered to be potential health
hazards. To view or print a Material Safety Data Sheet (MSDS) for these reagents
go to www.lamotte.com. To obtain a printed copy, contact LaMotte by e-mail,
phone or fax.
Cobalt rarely occurs in natural water. It is used in the manufacture of alloys
to increase corrosion resistance and strength. It is found in wastewaters as a
corrosion by-product.
APPLICATION:
Industrial wastewater.
RANGE:
0.00–2.00 Cobalt
MDL:
0.02 ppm
METHOD:
PAN (1-(2-Pyridylazo)-2-Naphthol) forms a greenish
complex with Cobalt (Co
+2
) at a pH of 5.
SAMPLE HANDLING &
PRESERVATION:
Store samples in acid-washed plastic bottles. Adjust
pH to less than 2 with nitric acid. Adjust sample pH to 5
before testing.
INTERFERENCES:
Iron (+2) and high concentrations of heavy metals.
SMART Spectro Test Procedures 2.11
COBALT
Test P
rocedures