PROCEDURE

Use universal sample holder.

NOTE: Place Dispenser Cap (0692) on *Mixed Acid Reagent (V-6278). Save this

cap for refi ll reagents.

1. Press and hold ON button until spectrophotometer turns on.
2. Scroll to and select PROGRAMMED TESTS.
3. Scroll to and select ALL TESTS (or another sequence containing

67 Nitrite-N LR) from TESTING MENU.

4. Scroll to and select 67 Nitrite-N LR from menu.
5. Rinse a clean tube (0290) with sample water. Fill to the 10 mL line with

sample.

6. Insert tube into chamber, close lid and select SCAN BLANK.
7. Remove tube from Spectro and pour off 5 mL into a graduated cylinder or

similar. Discard the remaining sample.

8. Pour the 5 mL sample from the graduated cylinder into the colorimeter tube.

Use graduated cylinder or similar to measure 5 mL of *Mixed Acid Reagent (V-

6278) and add to tube. Cap and mix.

9. Use the 0.1 g spoon (0699) to add two measures of *Color Developing

Reagent (V-6281). Cap and mix by gently inverting for 1 minute. Wait 5 minutes

for maximum color development.

10. At the end of the 5 minute waiting period, mix, insert tube into chamber, close

lid and select SCAN SAMPLE. Record result.

11. Press OFF button to turn spectrophotometer off or press EXIT button to exit to

a previous menu or make another menu selection.

NOTE: To convert nitrite-nitrogen (NO

2

–N) results to ppm nitrite (NO

2

), multiply

results by 3.3.

NITRITE-NITROGEN, Low Range

SMART Spectro Test Procedures 2.11