beautypg.com

Procedure – LaMotte SMART Spectro Spectrophotometer User Manual

Page 179

background image

PROCEDURE

Use universal sample holder.

1. Press and hold ON button until spectrophotometer turns on.
2. Scroll to and select PROGRAMMED TESTS.
3. Scroll to and select ALL TESTS (or another sequence containing

54 Lead) from TESTING MENU.

4. Scroll to and select 54 Lead from menu.
5. Rinse a tube (0290) with sample water. Fill to 10 mL with sample.
6. Insert the tube into chamber, close lid and select SCAN BLANK.
7. Remove the tube from Spectro. Empty the tube. Use the Syringe (0807) to add

5mL of sample to the tube.

8. Add 5 mL *Ammonium Chloride Buffer (4032) to fi ll the tube to the 10 mL line.

Swirl to mix.

9. Add 3 drops *Sodium Cyanide, 10% (6565). Swirl to mix.
10. Use the 0.5 mL pipet (0353) to add 0.5 mL *PAR Indicator (4033). Swirl to

mix.

11. Use the 0.5 mL pipet (0353) to add 0.5 mL Stabilizing Reagent (4022). Cap

and mix.

12. Insert tube into chamber, close lid and select SCAN SAMPLE. Record result

in ppm as Reading A.

13. Remove tube from Spectro. Add 3 drops *DDC Reagent (4034). Cap and

mix.

14. Insert tube into chamber, close lid and select SCAN SAMPLE. Record result

in ppm as Reading B.

15. ppm Lead = Reading A–Reading B
16. Press OFF button to turn the spectrophotometer off or press EXIT button to

exit to a previous menu or make another menu selection.

NOTE: For best possible results, a reagent blank should be determined to account

for any contribution to the test result by the reagent system. To determine the

reagent blank, follow the above test procedure to scan a distilled or deionized

water blank. Then follow the above procedure to perform the test on a distilled or

deionized water sample. This test result is the reagent blank. Subtract the reagent

blank from all subsequent test results of unknown samples. It is necessary to

determine the reagent blank only when a new lot number of reagents is obtained.

LEAD

SMART Spectro Test Procedures 2.11

Test P

rocedures