beautypg.com

Procedure–free chlorine, Procedure–combined chlorine – LaMotte SMART Spectro Spectrophotometer User Manual

Page 90

background image

PROCEDURE–FREE CHLORINE

Use universal sample holder.

1. Press and hold ON button until spectrophotometer turns on.
2. Scroll to and select PROGRAMMED TESTS.
3. Scroll to and select ALL TESTS (or another sequence containing

15 Chlorine) from TESTING MENU.

4. Scroll to and select 15 Chlorine from menu.
5. Rinse a clean tube (0290) with sample water. Fill to the 10 mL line with

sample.

6. Insert tube into chamber, close lid and select SCAN BLANK.
7. Remove tube from Spectro and pour off all but a suffi cient amount of sample

water to cover a tablet. Add one *Chlorine DPD #1 Instrument Grade Tablet

(6903A). Crush tablet with a tablet crusher (0175), then add sample water

until tube is fi lled to 10 mL line. Cap tube and shake until tablet has dissolved.

Solution will turn pink if free chlorine is present. Wait 15 seconds, but no longer

than 30 seconds. Mix.

8. Insert tube into chamber, close lid and select SCAN SAMPLE.

PROCEDURE–COMBINED CHLORINE

Use universal sample holder.

9. Add one *Chlorine DPD #3 Instrument Grade Tablet (6197A) to sample from

Step 8 above. Crush tablet with tablet crusher (0175). Cap tube and shake until

tablet dissolves. An increase in color represents combined chlorine.

NOTE: For wastewater samples, Standard Methods for the Examination

of Water and Wastewater recommends waiting 2 minutes for full color

development.

10. Insert sample into chamber, close lid and select SCAN SAMPLE. Record

result as Total Chlorine.

11. Subtract free chlorine reading from total chlorine reading to obtain

concentration of combined chlorine.

12. Press the OFF button to turn off the spectrophotometer or press the EXIT

button to exit to a previous menu or make another menu selection.

SMART Spectro Test Procedures 2.11

CHLORINE, BROMINE, IODINE

Test P

rocedures