beautypg.com

Procedure – LaMotte SMART Spectro Spectrophotometer User Manual

Page 152

background image

PROCEDURE

Use 10 mm square cell adapter.

1. Press and hold ON button until spectrophotometer turns on.
2. Scroll to and select PROGRAMMED TESTS.
3. Scroll to and select ALL TESTS (or another sequence containing

37 Cyanuric-UDV) from TESTING MENU.

4. Scroll to and select 37 Cyanuric-UDV from menu.
5. Rinse a clean vial (0156) with sample water.
6. Use the syringe (1184) to add 3 mL of sample to the vial.
7. Insert the vial into chamber, close lid and select SCAN BLANK.
8. Remove vial from Spectro.
9. Use the syringe (1184) to add 3 mL of sample to a Cyanuric Acid UDV vial

(4313).

10. Invert vial 3 times to mix.
11. Wait 2 minutes.
12. Invert vial 3 times to mix.

NOTE: If powder residue remains in the bottom of the vial after inverting, or air

bubbles forms, invert vial once more and tap bottom of vial sharply once or

twice to dislodge powder or bubbles. Mix.

13. Insert tube into chamber, close lid and select SCAN SAMPLE. Record

result.

14. Press OFF button to turn the spectrophotometer off or press EXIT button to

exit to a previous menu or make another menu selection.

NOTES: For best possible results, a reagent blank should be determined to

account for any contribution to the test result by the reagent system. To determine

the reagent blank, follow the above test procedure to scan a distilled or deionized

water blank. Then follow the above procedure to perform the test on a distilled or

deionized water sample. This test result is the reagent blank. Subtract the reagent

blank from all subsequent test results of unknown samples. It is necessary to

determine the reagent blank only when a new lot number of reagents is obtained.

UDVs from opened pouches should be used promptly. Store unused vials from

opened pouches in the Foil Storage Bag (9467) to extend the shelf life of the

reagent. Generally, UDVs stored in the bag should be used within 10 days if the

humidity is less than 50% and within 5 days if humidity is greater than 50%. The Foil

Storage Bag contains a desiccant pack

SMART Spectro Test Procedures 2.11

CYANURIC ACID, UDV

Test P

rocedures