Procedure – LaMotte SMART Spectro Spectrophotometer User Manual
Page 171
PROCEDURE
Use universal sample holder.
1. Press and hold ON button until spectrophotometer turns on.
2. Scroll to and select PROGRAMMED TESTS.
3. Scroll to and select ALL TESTS (or another sequence containing
51 Iron Bipyr) from TESTING MENU.
4. Scroll to and select 51 Iron Bipyr from menu.
5. Rinse a clean tube (0290) with sample water. Fill to the 10 mL line with
sample.
6. Insert tube into chamber, close lid and select SCAN BLANK.
7. Remove the tube from Spectro. Use the 0.5 mL pipet (0353) to add on
measure of *Iron Reagent #1 (V-4450). Cap and mix.
8. Use the 0.1 g spoon (0699) to add 0.1 g of *Iron Reagent #2 Powder (V4451).
Cap and shake vigorously for 30 seconds. Wait three minutes for maximum
color development.
9. At the end of 3 minute waiting period, do not mix. Insert tube into chamber,
close lid and select SCAN SAMPLE. Record result.
10. Press OFF button to turn spectrophotometer off or press EXIT button to exit to
a previous menu or make another menu selection.
NOTE: For best possible results, a reagent blank should be determined to account
for any contribution to the test result by the reagent system. To determine the
reagent blank, follow the above test procedure to scan a distilled or deionized
water blank. Then follow the above procedure to perform the test on a distilled or
deionized water sample. This test result is the reagent blank. Subtract the reagent
blank from all subsequent test results of unknown samples. It is necessary to
determine the reagent blank only when a new lot number of reagents is obtained.
IRON, Bipyridyl
SMART Spectro Test Procedures 2.11
Test P
rocedures