15 current sample data, key Page 98
Page 98

2.15 Current sample data, key
751 GPD Titrino
95
2.15 Current sample data, key
The key
current sample data. The contents of this key change
when the silo memory is switched on, see page 97.
Instead of entering the current sample data with
automatically after start of determinations.
Configuration: , ">preselections".
Current sample data can be entered live. For working
with the silo memory see page 96.
id
#1...3 or C21...C23, sample identifications:
The sample identifications can also be used as sample
specific calculation variables C21...C23.
smpl size:
Sample size.
The sample size can be monitored, see e.g. page 19.
The limits are then displayed in this window.
smpl unit:
Unit of the sample size.
The display texts of the Titrino are shown to the left.
The values are the default values.
smpl data
id
#1 or C21
id
#2 or C22
id
#3 or C23
smpl size 1.0 g
smpl unit: g
Sample data
Sample identification 1...3 or sample specific operand
C21...C23 (up to 8 characters).
Sample identifications or sample specific operands
can be entered using the keypad, via a balance with a
special input device or via barcode reader.
Sample size (6-digit number: ±X.XXXXX)
Entry using keypad, via balance or via barcode reader.
Unit of sample size (g, mg, ml, ul, pc, no unit or up to 5
characters)
Select unit with <
←
/
→
>.
SMPL
DATA
smpl data
id
#1 or C21
id
#2 or C22
id
#3 or C23
smpl size 1.0 g
smpl unit: g