Metrohm 751 GPD Titrino User Manual
Page 146

3.2 Remote control commands
751 GPD Titrino
143
3.2.2 Description of the remote control commands
3.2.2.
3.2.2.1.
1.
M
Mode
$G, $S, $H, $C
Start and stop ($G, $S) or hold of the current method (3.2.2.3) with $H and
continue with $C.
$G also serves to continue after inquiries of identifications and sample size af-
ter the start (see 3.2.2.26) as well as after inquiries of calibration temperature
and pH values of buffers (see 3.2.2.68 and 3.2.2.69).
3.2.2.
3.2.2.2.
2.
M
Mode.Q
.QuickMeas
$G, $S
Start and stop of a measurement in the basic mode with the parameters
(measured quantity, measuring input) of the current method. Corresponds to
the
&Mode.Parameter, see 3.2.2.76 and 3.2.2.56.
With an ongoing measurement, the current mode can be started. This stops the
measurement automatically.
3.2.2.
3.2.2.3.
3.
M
Mode.S
.Select
DET
DET, MET, SET, KFT, STAT, DOS,
DOC, MEAS, CAL, TIP
M
Mode.D
.DETQuantity
pH
pH, U, Ipol, Upol
M
Mode.M
.METQuantity
pH
pH, U, Ipol, Upol
M
Mode.SET
.SETQuantity
pH
pH, U, Ipol, Upol
M
Mode.MEA
.MEASQuantity
pH
pH, U, Ipol, Upol, T
M
Mode.K
.KFTQuantity
Ipol
Ipol, Upol
M
Mode.ST
.STATQuantity
pH
pH, U, Ipol, Upol
M
Mode.DO
.DOCQuantity
pH
pH, U, Ipol, Upol
Selection of the standard mode. Mode and the measured quantity belong to the
complete selection.
If a method is selected from the method memory, the nodes &Mode.Select and
&Mode.XXXQuantity are overwritten with mode and measured quantity of the
corresponding user method.
3.2.2.
3.2.2.4.
4.
M
Mode.N
.Name
read only
Name of the current method in the working memory. $Q sends 8 ASCII char-
acters. Standard methods carry the name ********. The node can be set read
+ write, see 3.2.2.134.
3.2.2.
3.2.2.5.
5.
M
Mode.Def.F
.Def.Formulas.1.F
.1.Formula
EPX, CXX, RSX, +, -, *, /, (, )
M
Mode.Def.F
.Def.Formulas.1.T
.1.TextRS
up to 8 ASCII characters
M
Mode.Def.F
.Def.Formulas.1.D
.1.Decimal
0...22...5
M
Mode.Def.F
.Def.Formulas.1.U
.1.Unit
up to 6 ASCII characters
M
Mode.D
.Def.F
.Formulas.1.L
.1.Limits
ON, OFF
OFF
M
Mode.D
.Def.F
.Formulas.1.Lo
.1.LoLim
00...
±
999 999
M
Mode.D
.Def.F
.Formulas.1.Up
.1.UpLim
00...
±
999 999
M
Mode.D
.Def.F
.Formulas.1.O
.1.Output
active, pulse, OFF
OFF
M
Mode.D
.Def.F
.Formulas.2.F
.2.Formula
etc. up to .9
.9