Metrohm 751 GPD Titrino User Manual
Page 153
3.2. Remote control commands
751 GPD Titrino
150
Parameter for SET, KFT: Pause time in s. Runs before dosing the start volume.
3.2.2.
3.2.2.34.
34. M
Mode.P
.Parameter.T
.TitrPara.E
.ExtrT
00...999 999
Parameter for SET, KFT: Extraction time in s.
3.2.2.
3.2.2.35.
35. M
Mode.P
.Parameter.T
.TitrPara.TD
.TDelta
1...22...999 999
Parameter for SET, KFT, STAT: Time interval in s for the entry of a measure-
ment point in the list of measured points.
3.2.2.
3.2.2.36.
36. M
Mode.P
.Parameter.P
.Presel.C
.Cond
ON, OFF
OFF
M
Mode.P
.Parameter.P
.Presel.DDriftDisp
ON
ON, OFF
M
Mode.P
.Parameter.P
.Presel.DC
DCor.T
.Type
auto, man., OFF
OFF
M
Mode.P
.Parameter.P
.Presel.DC
DCor.V
.Value
0.0
0.0...99.9
Parameters for SET, KFT:
.Cond: Conditioning ON/OFF
.DriftDisp: Drift display during conditioning ON/OFF.
.DCor.Type: Type of drift take-over for the drift correction. auto: Take-over of
the drift value at start.
.DCor.Value: Drift value for the manual drift correction.
3.2.2.
3.2.2.37.
37. M
Mode.P
.Parameter.C
.CtrlPara.E
.EP
pH:
0...
±
20.00, OFF
OFF
U, Ipol: 0...
±
2000, OFF
OFF
Upol: 0...
±
200.0, OFF
OFF
M
Mode.P
.Parameter.C
.CtrlPara.U
.UnitEp
read only
Parameters for KFT, STAT: Setting of the EP, resp. Control point as a pH value,
in mV (with U and Ipol) or uA (with Upol). The corresponding unit can be read
with .UnitEP.
With KFT the standard value for Ipol is 250 mV, for Upol 25 uA.
3.2.2.
3.2.2.38.
38. M
Mode.P
.Parameter.C
.CtrlPara.D
.Dyn
pH: 0.01...20.00, OFF
OFF
U, Ipol: 1...2000, OFF
OFF
Upol: 0.1...200.0, OFF
OFF
M
Mode.P
.Parameter.C
.CtrlPara.UnitD
.UnitDyn
read only
M
Mode.P
.Parameter.C
.CtrlPara.M
.MaxRate
0.01...10
10...150, max.
M
Mode.P
.Parameter.C
.CtrlPara.Mi
.MinIncr
0.1...9.9, min.
min.
Parameters for KFT, STAT: Control parameters.
.Dyn: Dynamics (control range) in pH, mV (with U and Ipol) or uA (with Upol).
The corresponding unit can be read with .UnitDyn. Standard with
KFT Ipol 100, with KFT Upol 10.
.MaxRate: Maximum allowed titration rate in ml/min. Max. means maximum
possible rate with mounted Exchange Unit.
.MinIncr: Parameter for KFT. Minimum volume increment in ul.
3.2.2.
3.2.2.39.
39. M
Mode.P
.Parameter.P
.Presel.O
.Oven
COM1, COM2, no
no
Parameter for KFT.