Metrohm 751 GPD Titrino User Manual
Page 157

3.2. Remote control commands
751 GPD Titrino
154
3.2.2.
3.2.2.56.
56. M
Mode.P
.Parameter.M
.Monitoring.M
.MeasVal.M
.MeasModepH,U,Ipol,Upol,OFF
OFF
M
Mode.P
.Parameter.M
.Monitoring.M
.MeasVal.MeasI
.MeasInput
11, 2, diff.
M
Mode.P
.Parameter.M
.Monitoring.M
.MeasVal.P
.PCurrent
0...11...
±
127
M
Mode.P
.Parameter.M
.Monitoring.M
.MeasVal.PV
.PVoltage
0...400
400...
±
1270
M
Mode.P
.Parameter.M
.Monitoring.M
.MeasVal.Po
.PolElectrTest
ON, OFF
OFF
M
Mode.P
.Parameter.M
.Monitoring.M
.MeasVal.pH.L
.pH.LowLim
00...
±
20.00
M
Mode.P
.Parameter.M
.Monitoring.M
.MeasVal.pH.U
.pH.UpLim
0...
±
20.00
20.00
M
Mode.P
.Parameter.M
.Monitoring.M
.MeasVal.U.L
.U.LowLim
00...
±
2000
M
Mode.P
.Parameter.M
.Monitoring.M
.MeasVal.U.U
.U.UpLim
0...
±
2000
2000
M
Mode.P
.Parameter.M
.Monitoring.M
.MeasVal.I.Ipol.L
.LowLim
00...
±
2000
M
Mode.P
.Parameter.M
.Monitoring.M
.MeasVal.I.Ipol.U
.UpLim
0...
±
2000
2000
M
Mode.P
.Parameter.M
.Monitoring.M
.MeasVal.Up
.Upol.L
.LowLim
00...
±
200.0
M
Mode.P
.Parameter.M
.Monitoring.M
.MeasVal.Up
.Upol.U
.UpLim
0...
±
200.0
200.0
M
Mode.P
.Parameter.M
.Monitoring.M
.MeasVal.A
.Action end, hold, wait, none
none
Parameters for DOS: Selection of the measured quantity for measured value
monitoring. Measured values (pH, U and I) are entered in the list of measured
points only when the monitoring is active.
Selection of the measurement input (MeasInput) applies with measured quanti-
ties pH and U. diff. means differential amplifier, see page 215.
With Ipol, the requests for the polarization current in uA (Ipol) and .PolElectrTest
apply.
With Upol, the request for the polarization voltage in mV (Upol) applies. Entry in
steps of 10 mV. .PolElectrTest also applies. If the test for polarized electrodes
is switched on (ON), it is performed on the change from the inactive basic
mode to an active mode.
On limit value violation, an action can be selected:
stop: Abort of the determination.
hold: Interrupt dosing until the sequence is continued by manual switching (&m
$C).
wait: Interrupt dosing until the limit value is no longer violated, then continue
automatically.
3.2.2.
3.2.2.57.
57. M
Mode.P
.Parameter.D
.DosPara.B
.BeginMeas
pH:
0...
±
20.00, init
init
U, Ipol: 0...
±
2000, init
init
Upol:
0...
±
200.0, init
init
M
Mode.P
.Parameter.D
.DosPara.E
.EndMeas
as with .Begin, but OFF instead of init
M
Mode.P
.Parameter.D
.DosPara.U
.UnitMeas
read only
Parameters for DOS: Setting of ramp start and ramp end. init means ramp start
with default measured value.
3.2.2.
3.2.2.58.
58. M
Mode.P
.Parameter.D
.DosPara.S
.SweepTime
0...300
300...999 999
Sweep time for the ramp in s.
3.2.2.
3.2.2.59.
59. M
Mode.P
.Parameter.D
.DosPara.D
.Dyn
pH: 0.01...0.25
0.25...20.00, OFF
U:
1...15
15...2000, OFF
Ipol:
1...66...2000, OFF