3.2. Remote control commands
751 GPD Titrino
128
*Parameter Tree part "Parameters for MEAS"
Ã
.
Measuring
Measuring parameters
³ Ã
.SignalDrift
Drift for meas.value acquisition
depends on meas.quant. 3.2.2.63.
³ Ã
.UnitSigDrift
Unit of measured value drift
read only
ditto
³ Ã
.EquTime
Equilibrium time
0...9999, OFF
ditto
³ Ã
.MeasInput
Measuring input
1, 2, diff.
3.2.2.64.
³ Ã
.Ipol
Polarization current
0...1...
±
127
ditto
³ Ã
.Upol
Polarization voltage
0...400...
±
1270
ditto
³ Ã
.PolElectrTest
Test for polarized electrodes
ON, OFF
ditto
³ Ã
.Temp Titration temperature
-170.0...25.0...500.0
3.2.2.65.
³ Ã
.TDelta
Time interv.for meas.acquisition
1...2...999 999
3.2.2.66.
³ ³
Ã
.
Statistics Statistics
³ Ã
.Status
Status of statistics calculation
ON, OFF
3.2.2.24.
³ Ã
.MeanN
No. of individual determinations
2...20
ditto
³ Ã
.ResTab
Result table
³ ³ Ã
.Select
original,delete n,delete all
ditto
³ ³ Ã
.DelN
Deletion of individual results
1...20
ditto
³ ³
Ã
.
Presel
Preselections
³ Ã
.IReq
Request of Id's after start
id1, id1&2, all, OFF
3.2.2.26.
³ Ã
.SReq
Request of sample size after start
value, unit, all, OFF
ditto
³ Ã
.LimSmplSize
Limits for sample size
3.2.2.27
³ ³ Ã
.Status
Status of limit control
ON, OFF
ditto
³ ³ Ã
.LoLim
Lower limit
0.0...999 999
ditto
³ ³ Ã
.UpLim
Upper limit
0.0...999 999
ditto
³ Ã
.ActPulse
Output of a pulse
ON, OFF
3.2.2.28.
*Parameter Tree part "Parameters for CAL"
Ã
.
Calibration
Calibration parameters
³ Ã
.MeasInput
Measuring input
1, 2, diff.
3.2.2.67.
³ Ã
.CalTemp
Calibration temperature
-20.0...25.0...120.0
3.2.2.68.
³ Ã
.Buffer
³ ³ Г
.1
³ ³ ³ Г
.Value
pH value of buffer 1
0...7.00...
±
20.00
3.2.2.69.
³ ³ Г
.2
³ ³ ³ Г
.Value
pH value of buffer 2
0...4.00...
±
20.00, OFF
ditto
³ ³ Ã
:
up to 9 buffers
³ Ã
.SignalDrift
Drift for meas.value acquisition
depends on meas.quant. 3.2.2.70.
³ Ã
.EquTime
Equilibrium time
0...26...9999, OFF
ditto
³ Ã
.ElectrodeId
Electrode identification
8 ASCII char.
3.2.2.71.
³ Ã
.SmplChanger
Calibration on a Titrino
ON, OFF
3.2.2.72.
³ Ã
.ActPulse
Output of a pulse
first, all, OFF
3.2.2.73.
³ ³
Ã
.
Statistics Statistics
³ Ã
.Status
Status of statistics calculation
ON, OFF
3.2.2.24.
³ Ã
.MeanN
No.of individual determinations
2...20
ditto
³ Ã
.ResTab
Result table
³ ³ Ã
.Select
original,delete n,delete all
ditto
³ ³ Ã
.DelN
Deletion of individual results
1...20
ditto