Metrohm 751 GPD Titrino User Manual
Page 152

3.2 Remote control commands
751 GPD Titrino
149
3.2.2.
3.2.2.28.
28. M
Mode.P
.Parameter.P
.Presel.A
.ActPuls
ON, OFF
OFF
for SET, KFT: first, all, cond., OFF
OFF
Output of a pulse on the I/O line "Activate", see page 224.
3.2.2.
3.2.2.29.
29. M
Mode.P
.Parameter.S
.SET1.E
.EP
pH:
0...
±
20.00, OFF
OFF
U, Ipol: 0...
±
2000, OFF
OFF
Upol:
0...
±
200.0, OFF
OFF
M
Mode.P
.Parameter.S
.SET1.U
.UnitEp
read only
Parameters for SET: Setting the 1st endpoint as pH value, in mV (with U and
Ipol) resp. uA (with Upol). The corresponding unit can be read with .UnitEP.
If the value is on "OFF", no further nodes will appear from SET1.
3.2.2.
3.2.2.30.
30. M
Mode.P
.Parameter.S
.SET1.D
.Dyn
pH: 0.01...20.00, OFF
OFF
U, Ipol: 1...2000, OFF
OFF
Upol: 0.1...200.0, OFF
OFF
M
Mode.P
.Parameter.S
.SET1.UnitD
.UnitDyn
read only
M
Mode.P
.Parameter.S
.SET1.M
.MaxRate
0.01...10
10...150, max.
M
Mode.P
.Parameter.S
.SET1.Mi
.MinRate
0.01...25.0
25.0...999.9
Parameters for SET: Control parameters, see page 31.
.Dyn:
Dynamics, control range in pH, mV (with U and Ipol) or uA (with
Upol). The corresponding unit can be read with .UnitDyn.
.MaxRate: Maximum allowed titration rate in mL/min. Max. means maximum
possible rate with the Exchange Unit in current use.
.MinRate: Minimum titration rate in ul/min.
3.2.2.
3.2.2.31.
31. M
Mode.P
.Parameter.S
.SET1.S
.Stop.T
.Type
drift
drift, time
M
Mode.P
.Parameter.S
.SET1.S
.Stop.D
.Drift
1...20
20...999
M
Mode.P
.Parameter.S
.SET1.S
.Stop.Ti
.Time
0...10
10...999, inf
M
Mode.P
.Parameter.S
.SET1.S
.Stop.S
.StopT
0...99 999, OFF
OFF
Parameters for SET: Type and size of the stop criterion of the titration.
.Type: Type of stop criterion after stop drift or switch-off delay time.
.Drift: Stop drift in ul/min. Applies when "drift" has been selected.
.Time: Switch-off delay time in s. Applies when "time" has been selected. "inf"
means infinite.
.StopT: Stop time in s. Applies when "time" has been selected and the value of
.Time is set to "inf.".
3.2.2.
3.2.2.32.
32. M
Mode.P
.Parameter.T
.TitrPara.D
.Direction
+, -, auto
auto
Parameters for SET, KFT, STAT: Titration direction.
"auto" means the titration direction is determined automatically by the instru-
ment. If 2 EP's have been set in a SET titration, the titration direction is given by
the two EP's. The entry of the titration direction is then invalid.
3.2.2.
3.2.2.33.
33. M
Mode.P
.Parameter.T
.TitrPara.X
.XPause
00...999 999